An Inventory of Current Available Ultrasound Devices for Dental Use
RESEARCH ARTICLE

An Inventory of Current Available Ultrasound Devices for Dental Use

The Open Dentistry Journal 31 Jul 2015 RESEARCH ARTICLE DOI: 10.2174/1874210601509010319

Abstract

Objective : The first aim of the study was to make an inventory of the currently available ultrasound imaging devices on the market. The second aim was to identify portable ultrasound devices which could be used in general dental offices to image salivary glands, masticatory muscles and lymph nodes. Material and Methods : A list of ultrasound devices was collected from 11 international and American manufacturers using company websites and ultrasound retailer’s websites using the search engine Google®. Results : A total of 86 ultrasound machines from 11 different manufacturers were identified, of which only 14 devices are portable. Transducers are usually transferrable between devices of the same brand, and the non-portable machines have multiple transducer ports (three or four), while portable machines typically have only one or two ports. Portable machine weights range from 10 to 20 pounds, and often resemble large laptops, which are convenient to carry and take up minimal space in clinics. Conclusion : Some manufacturers provide portable and non-portable ultrasound devices. The first ones could be valuable in dental settings, as they can easily be moved around in the office.

Keywords: Color doppler, dental, imaging, maxillofacial, radiology, sonography, ultrasound.

INTRODUCTION

Ultrasonography is an imaging technique for visualizing organs and soft tissues inside the body using reflective high frequency sounds waves. A transducer probe Fig. (1) sends high frequency oscillating pulses toward an area of interest such as soft tissues, tendons, blood vessels, and organs and the same probe receives echoes that bounce back from the tissues. The differences in echo impedance correspond to different types of tissues and tissue composition. A computer converts these echo differences to create real-time images of these tissues [1-3]. Images are used to detect and treat a wide variety of pathologies and lesions in the body.

Fig. (1).

A linear transducer probe with a variable frequency of 7 to 12 MHz.

Fig. (2).

Illustration of the color Doppler effect – image from the sternocleidomastoid muscle area, with the jugular vein and carotid artery being shown with the color Doppler (motion artefacts also cause some color Doppler effects).

Fig. (3).

Illustration of a transverse ultrasound image of the floor of the mouth (1. Anterior bellies of the digastric muscles, 2. Mylohyoid muscle, 3. Genioglossus and geniohyideus muscles).

Table 1.

List of ultrasound imaging devices that meet the study criteria.

Manufacturer Name Portable Weight (lbs) Size (h x w x d) (inches) Transducer Port(s) Transducer Name and Frequency (Mhz) Doppler Color (Triplex) Doppler 3-D Option 4-D Option
Acuson S2000 no 425 51-62 x 25 x 43 3 6-18: 18L6HD,
5-14: 14L5
PW, CW yes yes yes
Acuson Antares no 406 54-60 x 24 x 36 3 5-13: VFX13-5, VF13-5 PW, CW yes yes yes
Acuson X700 no 225 55-66 x 22 x 36 3 4-12: VF12-4,
5-10: VF10-5
PW, CW yes yes yes
Acuson X300 PE no 225 54-65 x 20 x 35 3 3-10: VF8-3,
5-10: VF10-5,
5-13: VF13-5
PW, CW yes yes yes
Aloka Noblus yes / / 1 5-10: UST-5413,
2-10: ASU-1010
PW, CW yes yes yes
Aloka Prosound Alpha 7 no / / 3 5-13: UST-5411, UST-5412,
3-13: UST-567,
7-13: UST-547
PW, CW yes yes yes
Aloka Prosound Alpha 6 no 154.3 51.2-61 x 16.5 x 27.6 3 5-13: UST-568, UST-5413, UST-5524, UST-5710,
7-15: UST-5542
PW, CW yes yes yes
Aloka Prosound Alpha 10 no 463 57-62 x 23 x 43 4 5-13: UST-5411, UST-5412, UST-5543,
3-13: UST-567
PW, CW yes yes yes
Aloka Prosound Alpha 5SX no 397 52-67 x 22 x 39 3 5-13: UST-5410, UST-567 PW, CW yes yes yes
Aloka


HI VISION Avius no / 56.3 x 17.7 x 27.6 3 5-18: L75,
6-14: L65,
5-13: L74M, LV74, 5-10: L53, L53L
PW, CW yes yes yes
Aloka HI VISION Preirus no / 53-61 x 17.7 x 35.4-43.7 3 5-18: L75,
6-14: L65,
5-13: L74M, LV74, 5-10: L53, L53L
PW, CW yes yes yes
Aloka Hi VISION Ascendus no / 53-61 x 21.7 x 35.4-43.7 4 5-18: L75,
6-14: L65,
5-13: L74M, LV74, 5-10: L53, L53L
PW, CW yes yes yes
ATL HDI 5000 no 440 62 x 28 x 43 3 5-12: L12-5 PW, CW yes yes no
ATL HDI 4000 no 223 59 x 27 x 44 3 5-12: L12-5 PW, CW yes yes yes
Esaote MyLab 30 Gold CV yes / / 2 6-18: LA453,
4-13: BL433 appleprobe, LA523, 3-11: LA332
PW, CW yes yes yes
Esaote MyLab 25 Gold yes / / 2 6-18: LA453,
4-13: BL433 appleprobe, LA523, 3-11: LA332
PW, CW yes yes yes
Esaote MyLab 40 Family no 140 58 x 21 x 31.5 3 15-22: SL3116,
6-18: LA435, SL3235,
4-13: LA523,
3-11: LA332
PW, CW yes yes yes
Esaote MyLab 50 Family no / / 3 15-22: SL3116,
6-18: LA435, SL3235,
4-13: LA523,
3-11: LA332
PW, CW yes yes yes
Esaote MyLab ClassC no / / 4 15-22: SL3116,
6-18: LA435, SL3235,
4-13: LA523,
3-11: LA332
PW, CW yes yes yes

Esaote

MyLab Seven

no

/

/

4

6-18: LA435,
4-13: SL1543, LA523,
3-11: AL2442

PW, CW

yes

Yes

yes
Esaote MyLab Twice no / / 4 6-18: LA453,
4-13: BL433 appleprobe, LA523, 3-11: LA332
PW, CW yes yes yes
GE Logiq e yes 10 2.5x14x12 1 8-16: 16L-RS,
5-13: 12L-RS,
4-12: 8L-RS
PW,CW yes yes no
GE Logiq 9 no 434 55-66 x 25 x 35 4 4.5-16: 4D16L,
5-13: M12L,
3.5-11: 4D10L
PW, CW yes yes yes
GE Logiq 7 no 496 57-63 x 24 x 39 4 5-13: M12L,
4.5-13: 12L,
3.5-11: 4D10L
PW, CW yes yes yes
GE Logiq P6 no 176 53.5-60 x 16.9 x 25.2 3 5-15: ML6-15,
3-11: 11L
PW, CW yes yes yes
GE Logiq P5 no 165 55.5 x 16.9 x 25.2 3 4.5-13: 12L,
3.5-11: 4D10L,
4-10: 10L
PW, CW yes yes yes
GE Logiq 5 Expert no 397 57 x 20 x 39 3 4.5-13: 12L,
4-10: 10L
PW, CW yes yes no
GE Logiq 5 Pro no 375 57 x 20 x 39 3 4.5-13: 12L,
4-10: 10L
PW, CW yes yes no
GE Logiq S6 no / / 3 5-13: M12L,
4.5-13: 12L,
3.5-11: 4D10L
PW, CW yes yes yes
GE Logiq S7 (2012) no / / 4 5-15: ML6-15,
4-12: 11L-D,
3-10: 9L-D
PW, CW yes yes yes
GE Voluson 730 Pro no 300 56 x 27 x 39 4 6-18: RSP6-16,
5-17: SP10-16,
3-11: SP6-12
PW, CW yes yes yes
GE Voluson 730 Expert no 300 56x27x39 4 6-18: RSP6-16,
5-17: SP10-16,
3-11: SP6-12,
5-13: M12L
PW, CW yes yes yes
GE Voluson E6 no 289 55 x 23 x 37 3 7-18: SP10-16-D,
6-18: RSP6-16-D
PW,CW yes yes yes
GE Voluson E8 Expert no 289 55 x 33 x 37 4 4-13: ML6-15-D,
4-14: RM14L,
6-18: RSP6-16-D,
7-18: SP10-16-D
PW, CW yes yes yes
GE Vivid E9 no 309 45-53 x 22 x 33 5 2-10: 9L-D,
5-12: 11L-D,
4.5-15: ML6-15-D
PW, CW yes yes yes
GE Vivid 7 Pro no 419 54 x 25 x 34 4 4-10: 10L,
4.5-13: 12L,
5-14: M12L
PW, CW yes yes yes
Mindray M7 yes 14.3 3 x 14.2 x 14 1 6-14: L14-6s, L14-6Ns,
5-10: 7L4s
PW, CW yes yes yes
Mindray M5 yes 13.2 3x14.2x14 1 8-12: 10L4,
5-10: 7L4s
PW, CW yes yes no
Mindray DC8 no 244.7 52.6 x 23 x 36.6 4 3.5-16: L14-6NE, L14-6WE,
3-13: L12-3E
PW, CW yes yes yes
Mindray DC7 no 265 53.1 x 19.7 x 33.5 4 6-14: L14-6,
4-11: L11-4,
4-10: 7L4A
PW, CW yes yes yes
Mindray DC6 no 291 53-62 x 20 x 34 4 8-12: 10L4,
5-10: 7L4A, 7L6
PW, CW yes yes no
Mindray DC3 no 202 48-62 x 18 x 29 4 8-12: 10L4,
5-10: 7L6, 7L4A
PW,CW yes yes yes
Mindray DC-T6 no 209 50.8-61.8 x 18.1 x 28.7 4 6-14: L14-6, L14-6N, 4-12: L12-4,
5-10: 7L4A
PW, CW yes yes yes
Mindray DC-N3 no 136.7 55.7-50.2 x 19.7 x 27.2-29.1 4 6-14: L16-6,
4-12: L12-4,
5-10: 7L4A
PW, CW yes yes yes
Philips CX50 yes 13.6 3 x 16 x 14 1 3-12: L12-3 PW, CW yes yes no
Philips CX30 yes 13.6 3 x 16 x 14 1 4-12: L12-4 PW, CW yes yes /
Philips Epiq 7 no 309 / 4 5-18: L18-5,
3-12: L12-3
PW, CW yes yes yes
Philips Epiq 5 no 229.3 / 4 5-18: L18-5,
3-12: L12-3
PW, CW yes yes yes
Philips iE33 no 345 55-64 x 22 x 43 3 3-11: L11-3 PW, CW yes yes yes
Philips iU22 no 345 58 x 21.5 x 40.5 3 5-17: L17-5,
5-12: L12-5
PW, CW yes yes yes
Philips HD15 no 220 54-61x22x44 2 5-12: L12-5,
3-12: L12-3
PW, CW yes yes yes
Philips HD 11 XE no 220 51-58 x 21 x 44 5 5-12: L12-5,
3-12: L12-3
PW, CW yes yes yes
Philips HD9 no 222 51 x 20 x 35 4 5-12: L12-5 PW, CW yes yes yes
Philips HD7 XE no 165 52-58 x 22 x 40 4 3-12: L12-3,
5-12: L12-5
PW, CW yes yes no
Philips HD6 no / / 4 3-12: L12-3 PW, CW yes yes no
Philips EnVisor no 220 51-58 x 22 x 41 3 3-12:L12-3,
5-12:L12-5
PW, CW yes yes no
Philips Sonos 7500 no 478 57 x 24.5 x 36 3 3-11: L11-3 PW, CW yes yes yes
Philips ClearVue 650 no 115 46 x 21 x 23 4 4-12: L12-4 PW, CW yes yes yes
Philips ClearVue 550 no 115 54-60 x 21 x 23 3 4-12: L12-4 PW, CW yes yes no
Philips ClearVue 350 no 115 52-58 x 21 x 23 3 4-12: L12-4 PW, CW yes yes no
SIUI Apogee 1200 Touch yes / / 2 *linear probes ranging from 5-14 PW, CW yes yes yes
SIUI Apogee 1200 Omni yes / / 2 *linear probes ranging from 5-14 PW, CW yes yes yes
SIUI Apogee 3800 Touch no / / 4 5-14: L8L38C PW, CW yes yes yes
SIUI Apogee 3800 Omni no / / 4 5-14: L8L38C PW, CW yes yes yes
SIUI Apogee 3500 Touch no / / 4 5-14: L8L38C PW,CW yes yes yes
SIUI Apogee 3500 Omni no / / 4 5-14: L8L38C PW,CW yes yes yes
SonoScape S9 yes 16.2 / 2 5-10: L741,
5-12: L742, L752
PW, CW yes yes yes
SonoScape S8 yes 20 / 2 5-10: L741,L743,
5-12: L742, 6-12: 10L1
PW, CW yes yes yes
SonoScape S6 yes 17.4 / 2 5-10: L741, L743,
5-12: L742,
6-12: 10L1
PW, CW yes yes no
SonoScape S2 yes / / 2 5-10: L741,
6-12: 10L1
PW, CW yes yes yes
SonoScape S40 no 331 59.7 x 27 x 39.3 5 5-10: L741,
5-12: L742, L752
PW, CW yes yes yes
SonoScape S30 no / / 4 5-10: L741, L743,
5-12: L742, L752,
6-12: 10L1
PW, CW yes yes yes
SonoScape S20 no / / 4 5-10: L741,L743,
5-12: L742, L752
PW, CW yes yes yes
SonoScape SSI-5000 no 198.4 52 x 21.7 x 32.1 3 5-10: L741,
5-12: L742, 6-12: 10L1
PW, CW yes yes yes
SonoScape SSI-6000 no / / 3 5-10: L741, L743,
5-12: L742,
6-12: 10L1
PW, CW yes yes yes
SonoScape SSI-8000 no / / 4 5-10: L741, L743,
5-12: L742,
6-12: 10L1
PW, CW yes yes yes
Toshiba Aplio 300 no 309 53-68 x 20 x 35 4 7-14: PLT-1204BT, PLT- 1202s,
6-12: PLT-805AT,
5-11: PLT-704SBT, PLT-704AT
PW,CW yes yes yes
Toshiba Aplio 400 no 309 54-70 x 35 4 7-14: PLT-1204BT, PLT- 1202s,
6-12: PLT-805AT,
5-11: PLT-704SBT, PLT-704AT
PW,CW yes yes yes
Toshiba Aplio 500 no 375 55-70 x 23 x 35 4 7-18: PLT-1204BX, 7-14: PLT-1204BT, PLT- 1202s,
6-12: PLT-805AT,
5-11: PLT-704SBT, PLT-704AT
PW,CW yes yes yes
Manufacturer Name Portable Weight (lbs) Size (h x w x d) (inches) Transducer Port(s) Transducer Name and Frequency (Mhz) Doppler Color (Triplex) Doppler 3-D Option 4-D Option
Toshiba Aplio XG no 419 56-71 x 23 x 45-48 3 7-18: PLT-1204BX, 7-14: PLT-1204BT, PLT- 1202s,
6-12: PLT-805AT,
5-11: PLT-704SBT, PLT-704AT
PW,CW yes Yes yes
Toshiba Xario 200 no 165 51-60 x 19 x 31 3 7-14: PLT-1204BT, PLT- 1202s,
6-12: PLT-805AT,
5-11: PLT-704SBT, PLT-704AT
PW,CW yes yes yes
Toshiba Xario 100 no 155 51-60 x 19 x 31 3 7-14: PLT-1204BT, PLT- 1202s,
6-12: PLT-805AT,
5-11: PLT-704SBT, PLT-704AT
PW,CW yes yes yes
Toshiba Artida no 441 66 x 23 x 50 3 5-11: PLT-704SBT PW,CW yes yes yes
Ultrasonix SonixSP no 130 57 x 21 x 24 / 5-14: L14-5/38, L14-5/38 GPS, L14-5W/60,
8-40: L40-8/12
PW, CW yes yes yes
Ultrasonix SonixTouch no / / / 5-14: L14-5/38, L14-5/38 GPS, L14-5W/60, 4DL14-5/38,
8-40: L40-8/12
PW, CW yes yes yes
Ultrasonix SonixMDP no / / / 5-14: L14-5/38, L14-5/38 GPS, L14-5W/60, 4DL14-5/38,
8-40: L40-8/12
PW, CW yes yes yes
Table 2.

List of portable ultrasound imaging devices that meet the aims of this paper.

Manufacturer Name Portable Weight
(lbs)
Size
(h x w x d) (inches)
Transducer Port(s) Transducer Name and Frequency (Mhz) Doppler Color (Triplex) Doppler 3-D Option 4-D
Option
Aloka Noblus yes / / 1 5-10: UST-5413,
2-10: ASU-1010
PW, CW yes yes yes
Esaote MyLab 30 Gold CV yes / / 2 6-18: LA453,
4-13: BL433 appleprobe, LA523,
3-11: LA332
PW, CW yes yes yes
Esaote MyLab 25 Gold yes / / 2 6-18: LA453,
4-13: BL433 appleprobe, LA523,
3-11: LA332
PW, CW yes yes yes
GE Logiq e yes 10 2.5x14x12 1 8-16: 16L-RS,
5-13: 12L-RS,
4-12: 8L-RS
PW,CW yes yes no
Mindray M7 yes 14.3 3 x 14.2 x 14 1 6-14: L14-6s, L14-6Ns,
5-10: 7L4s
PW, CW yes yes yes
Mindray M5 yes 13.2 3x14.2x14 1 8-12: 10L4,
5-10: 7L4s
PW, CW yes yes no
Philips CX50 yes 13.6 3 x 16 x 14 1 3-12: L12-3 PW, CW yes yes no
Philips CX30 yes 13.6 3 x 16 x 14 1 4-12: L12-4 PW, CW yes yes /
SIUI Apogee 1200 Touch yes / / 2 *linear probes ranging from 5-14 PW, CW yes yes yes
SIUI Apogee 1200 Omni yes / / 2 *linear probes ranging from 5-14 PW, CW yes yes yes
SonoScape S9 yes 16.2 / 2 5-10: L741,
5-12: L742, L752
PW, CW yes yes yes
SonoScape S8 yes 20 / 2 5-10: L741,L743,
5-12: L742,
6-12: 10L1
PW, CW yes yes yes
SonoScape S6 yes 17.4 / 2 5-10: L741, L743,
5-12: L742,
6-12: 10L1
PW, CW yes yes no
SonoScape S2 yes / / 2 5-10: L741,
6-12: 10L1
PW, CW yes yes yes

Current use of ultrasound includes imaging unborn fetuses in pregnant women, measuring blood flow in heart valves and blood vessels in the body Fig. (2) and scanning abnormal tissues such as lumps in multiple parts of the body Fig. (3). The main advantages and differences of ultrasound over other imaging techniques such as X-ray imaging, is that it is not using potentially harmful ionizing radiation, and that it is of minimal discomfort for the patient and that it is relatively easy to use. In small dental offices, portable ultrasound devices offer clinicians flexibility in transferring the device from room to room and provide quick results using procedures much less expensive and invasive than other imaging procedures such as MRI (Magnetic Resonance Imaging) imaging, panoramic radiology, and computer tomography.

In contrast to audible sound which has frequencies between 20Hz to 20 kHz, ultrasonic waves have higher frequencies that are inaudible to the human ear. Audible sound waves have longer wavelengths and scatter more easily, making it not suitable for focused imaging of specific tissues or small areas of the body. Ultrasound for medical imaging has higher frequencies, typically ranging from 1 MHz to 15 MHz This attribute makes ultrasound appropriate for medical imaging as higher frequency waves reflect back at a faster rate and do not penetrate deep into the tissues, which enables better imaging of superficially located structures, in contrast to lower frequency ultrasound waves which will penetrate deeper into the soft tissues. Hard tissues, like bone or calculi obstructing a salivary duct, for instance, will reflect the ultrasound wave, but will not allow ultrasound penetration, resulting in a ‘black’ or hypoechoic shadow behind the bone or the obstruction. The latter is of diagnostic importance. In order for the ultrasound waves to be able to penetrate the patient’s soft tissues, firm contact between the transducer and the skin needs to be established. For the latter a gel is used as air is a bad medium for transmission of ultrasound waves [1-3].

The advantages of ultrasonography include lower costs relative to other soft tissue imaging procedures, more simple procedures and equipment are required, quick imaging and real time results, and it avoids certain limitations such as claustrophobia. With a competitive market for ultrasound devices, manufacturers are competing to increase the usability and applicability of their ultrasound machines in different fields of health care and customize their equipment with features such as high definition probes, Doppler effect, 3D and 4D imaging modalities. Besides that, they also aim at portability as it provides health care providers to move around in a clinical setting or even outside the clinical setting (e.g. nursing home) more easily [1-3].

There are currently 4 modes of ultrasonography imaging: A-mode, B-mode, M-mode (motion) and Doppler. A-mode uses one transducer to “scan a line through the body with the echoes plotted on screen as a function of depth” [4]. In B-mode, 2-D images are created using linear transducers that sends waves along one plane of the body. M-mode is a series of B-mode images taken in succession, creating moving images. Doppler mode measures the direction and velocity of blood flow, often used in cardiovascular imaging. The Doppler modality is useful in identifying regions of vascularity in lesions to perform fine needle aspiration biopsies (FNAB) and in avoiding areas of vascular tissue not intended to be treated [5]. Dangore-Khasbage et al. found high sensitivity in using color Doppler to contrast between reactive and metastatic cervical lymph nodes in oral cancer patients [6]. In studying 10 cases of intraoral cancerous growth in patients with oral and maxillary swelling of the neck, Joshi and Sudesh confirmed results from ultrasound correlated with the histopathology taken from incision biopsies and concluded ultrasound as “an adjunct to diagnosing oral and maxillary pathology” [7].

Many additional studies have supported the usefulness of ultrasonography in imaging tongue lesions and lymph node metastasis. Wakasugi-Sato et al. used B-mode ultrasonography for FNAB in diagnosing tumors in the neck region such as metastatic lymph nodes and masses in the salivary glands. A 7.5 MHz linear array transducer was used to measure the thickness and positions of tongue masses in their patients for diagnostic and therapeutic purposes. Kodama et al. used ultrasonography to measure carcinogenic tongue masses while Yuasa et al. used ultrasound imaging to detect cervical lymph nodes metastasis [8, 9].

The main goal of this report was to provide an overview of the currently available diagnostic ultrasound devices on the market. The growing interest in the field of oral and maxillofacial radiology to use ultrasonography for diagnosing lumps and bumps in the head and neck region (e.g. salivary glands, masticatory muscles, lymph nodes) makes this an interesting issue. Portability, small sized equipment, transducers that allow for relative superficial imaging and affordability play in this context an important role.

Currently, there are no known ultrasound devices made specifically for dental diagnosis and treatment. However, there is a growing market for ultrasound machines in clinical medicine that is continuing to see new advancements in functionality and technology. The use of ultrasonography in dental care has not been a major focus in the field of dentistry and dental education, nor in the field of ultrasound manufacturers, yet there exist convincing evidence that ultrasound has potential to be a useful tool in oral and maxillofacial care, which eliminates the use of ionizing radiation [10-14].

Aim

The first aim of the study was to make an inventory of the currently available ultrasound imaging devices on the market. The second aim was to identify portable ultrasound devices which could be used in general dental offices to image salivary glands, masticatory muscles and lymph nodes for instance.

MATERIALS AND METHOD

Because this was an attempt to obtain an inventory of different ultrasound machines on the market, PubMed was considered not to be the primary online search engine. A list of ultrasound devices was collected from 11 international and American manufacturers using company websites and ultrasound retailer’s websites using the search engine Google®. Inclusion factors of ultrasound machines studied included 3D/4D application, color Doppler function, pulsed wave and continuous wave Doppler function, high frequency linear transducers (at least of 10 MHz), and portability of the ultrasound machine. Doppler is based on the changes that occur in the frequencies of reflected ultrasound waves when the original wave transmitted from the probe hits a target in motion, such as red blood cells. Doppler is used to measure the direction and velocities of blood flow in the heart and blood vessels, and can detect the movement of soft tissues [15]. Pulsed wave Doppler (PW) pairs with 2-D imaging to measure the direction and velocity of blood flow in a specific area of blood vessels chosen by the transducer operator. The transducer sends a single pulse each time towards the tissue and receives a single corresponding reflecting pulse, registering where the reflection happened in the tissue and measuring the distance of the reflection back to the transducer. Operators can choose to image a sample volume to receive information about blood flow. Transducers using continuous wave Doppler continuously send pulses and receive reflecting frequencies, allowing for ultrasonic waves to hit more moving blood cells, capturing higher velocities of blood flow than pulsed wave Doppler [16]. Since pulses are continuously emitted on a line of the ultrasound beam, it is not possible to determine the velocity of a single area along the vessel being measured. Pulsed Doppler Wave (PW) is commonly used to access irregular blood flow or recognize the presence of blockages in arteries and veins. Color Doppler offer several advantages not provided by greyscale images. This function creates color coded images of the tissue, based on whether the blood is moving toward or away from the transducers. Color depth also provides information on relative speed of blood movement within the organs.

3-D imaging introduces technology that overcomes some of the challenges of 2-D imaging, which forces the image reader to imagine or visualize the area or tissue of interest in 3-D, lowering the efficiency and accuracy of diagnosis [17]. Since the transducer is manually controlled, it is difficult to detect the same area of interest if clinicians need to reevaluate a specific area [17]. Introducing a third plane makes this process more precise and less time consuming. 2-D imaging is less accurate than 3-D imaging in estimating the volume and size of masses and organs. While 3-D options have been widely built into current machines, newer models have a 4-D imaging options which are real time 3-D images.

Method Flowchart

Below is described how the search for the devices was conducted chronologically in Google®.

In Google® search engine type in: “ultrasound device brands”.

➜ List of manufacturers and machine names was subsequently found.

➜ Each specific manufacturer name was typed in again to find specific types of machines by that same manufacturer.

➜ Subsequently the specific name of the device followed by keywords such as “brochure” “specifications” “transducers” “transducer port” “dimensions” (ie. Aloka Noblus tranducers) was typed in to find technical information about the specific ultrasound device.

➜ If frequency of transducers was not found in the previous search, a separate search for “frequency” in Google® was performed”.

This procedure was performed for every manufacturer of ultrasound devices that could be identified through a Google® web search.

Websites used to conduct technical detail research were mostly ultrasound manufactures’ websites and online listings by secondary ultrasound retail companies. Websites were chosen based on their relevance and helpfulness at providing answers to researched questions about specific details of ultrasound devices such as transducer frequency. Websites providing brief overviews or pricing of devices that required website subscriptions or personal information were not considered for research. Websites that provided information on devices that deviated from information provided by other sources were not used.

RESULTS

A total of 86 ultrasound machines from 11 different manufacturers are included in the final list of devices that meet the research criterions (Table 1). All have the minimum inclusion features required for oral and maxillofacial applications. Only 14 devices are portable (Table 2). Transducers are usually transferrable between devices of the same brand, and the non-portable machines have multiple transducer ports (three or four), while portable machines typically have only one or two ports. One non-portable device, the Sonoscape S-40®, has five transducer ports. This brand also has the highest number of portable devices, namely four. Manufacturers Philips® and General Electric® appeared to have the greatest diversity in models. The weights of the portable ultrasound machines vary between ten to twenty pounds. Non-portable ultrasound machines weigh considerably more than portable machines and have significantly larger builds. Portable machines range from 10 to 20 pounds, and often resemble large laptops, which is convenient to carry and take up minimal space in clinics. The typical non-portable devices weigh anywhere from two hundred up to five hundred pounds. Non portable devices are typically 50 to 60 inches tall and 20 to 30 inches wide, making them a less feasible option for smaller offices.

DISCUSSION

The use of ultrasound is increasingly being incorporated in dental care technology. Mukhi and Mahindra and Pandley et al. supported the accuracy of ultrasonography imaging for diagnosing facial space infections (e.g. abscess, cellulitis) and for ultrasound guided fine needle aspiration (e.g. fluid aspiration from a fluid occupied space). Ultrasound images were effective in assessing the size and locations of abscess cavities, including predicting the quantity of pus in the cavities [18, 19].

Tikku et al. used color Doppler ultrasonography to study peripheral lesions blood flow to monitor patients after receiving endodontic surgery. They found that unlike conventional radiography, ultrasonography imaging could measure blood velocities in tissues, providing information about the healing stages of bone by assessing bone vascularity in addition to soft tissues at the operation site [20]. After surgery, the ultrasonography imaging showed increased blood flow, due to inflammation, but as healing occurred and inflammation decreased, blood velocities slowed. Color Doppler can distinguish between venous and arterial flow (the latter indicating healing) [20]. Color Doppler has significant potential in the future of dentistry by allowing clinicians to assess vascularity of maxillofacial lesions and bone tissues.

Lakshimi et al. evaluated ultrasound in detecting benign versus cervical lymph nodes using cytopathological evaluation as the gold standard. They used an Esaote MyLab-40® ultrasound machine and a 7.5 MHz linear transducer to make qualitative observations of cervical lymph nodes including lymph node size, shape, and nodal borders (sharp or smooth), hypoechoic or hyperechoic lymph node characteristics and the presence of necrotic tissues in the nodes. They found that examining the qualitatively results from ultrasonography had high sensitivity and specificity in differentiating the various stages of reactivity of lymph nodes [21].

Several recent publications have also emphasized on the applicability of ultrasound in maxillofacial diagnostics (e.g. temporomandibular joint, tongue, salivary glands and masticatory muscles), hence the urge for this publication to focus on the availability of ultrasound devices that can be used in a dental setting. It is obvious that in the near future ultrasonography will become part of the armamentarium of the dental professional or at least will become more custom amongst oral and maxillofacial radiologists. The latter is in some countries already the case (e.g. United Kingdom, Japan and Sweden). Experiments are being executed to even perform dental diagnostics with ultrasonography [22-26]. It is clear that ultrasonography is gaining more terrain and interest from manufacturers and clinicians in oral and maxillofacial diagnostics.

Manufacturers are constantly improving and increasing the usability of their machines by adding new features to keep their machines marketable next to their competition. Some ultrasound machines now come with touchscreen panels which may have hygienic advantages over raised-button panels. One disadvantage of machines with touch screen monitors is possible inaccuracy of reading images due to debris or finger smudges from clinicians touching the screen.

Portability and size is an important factor for dentists to consider, especially for practitioners with limited office space. Portable machines generally have less transducer ports than non-portable machines. The latter is probably less of a limiting factor for dental professionals as they will probably use one or two transducers only (e.g. one for intraoral and one for extraoral use). Many portable models from manufactures such as Philips® or Sonoscape® have carts, useful for dentists who want to switch to stationary clinical use and keep the portability feature of their devices. However, any cart will do, of course, for this purpose.

All manufactures in this study offer a variety of transducer probes such as convex, linear, phased array, endocavity and intra-operative probes. For the purpose of imaging the head and neck region, linear probes with rectangular and flat surfaces are most appropriate for placing against areas such as the jaw and neck. Most linear probes are compatible with multiple devices. Generally, portable devices offer transducers with slightly lower frequencies than non-portable devices. The most common types for portable machines are 5-10 MHz and 6-12 MHz probes, but there are exceptions such as the Esaote MyLab® portable series which offer probes up to 18 MHz The Ultrasonix Sonix® series offer the L40 probe that delivers the highest frequency found of all devices in this research at 40 MHz Other portable devices that offer high frequency transducers include the GE Logiq e® with the 16L-RS probe (8-16 MHz) and the Mindray M7® offering probes with 6-14 MHz imaging capabilities. While portable machines offer the convenience of smaller size, lower cost, and portability, clinicians looking for higher frequency imaging may find a limited number of portable devices that can meet the performance of a full-sized ultrasound machine.

All of the machines reviewed in this study have 3D option, and some also have 4D feature which is often referred to as “real-time 3D” which creates moving 3D images. For oral and maxillofacial applications, this feature might prove valuable and useful for temporomandibular joint diagnosis as visualizing the trajectory of the mandibular condyle may improve the clinician’s interpretation of the image and help diagnosing the patient’s problem. Also in salivary gland disease it may be useful to have a better visual of a gland in three dimensions.

Since the main objective of this research paper was to obtain an inventory of current ultrasound machines available on the market, the primary source used to research machines was manufacturers’ websites. Most manufacturers required subscriptions or customers to provide contact information in order to access additional specifications and information on machines. This limited the ability to obtain certain information such as dimensions and weight of many machines, since results are based on what manufacturers choose to advertise on their websites. For example, Siui® did not list specific transducers for their Apogee machines on their websites. Some information was obtained from secondary companies, therefore some specifications and details may deviate slightly from the primary manufacturers’ listings. This research was conducted from a consumer’s point of view in an attempt to quantify the manufacturers that provide portable ultrasound devices with high frequency transducers, which can be used in oral and maxillofacial radiology. It is clear that this report does not conclude anything with regard to image quality or scientific data related to these devices.

While this paper did not study the pricing of each machine, economical resource is an important factor for prospective consumers. Newer models and devices with more features or imaging options usually equate to increased prices. Portable devices that offer Color Doppler and 3D imaging cost more than black and white imaging devices. A university hospital or other larger corporations may have more resources to buy newer models, while clinicians with smaller budgets might consider an alternative and purchase refurbished machines, or consider buying more modest devices that still provide the essential functions for their imaging purposes. It is recommended that interested buyers conduct extensive research on ultrasound devices to find the product most appropriate for their needs.

CONCLUSION

There are several manufacturers that provide portable ultrasound devices with transducers in the higher MHz range, which can be useful for use in a dental setting. Portable devices these days offer the same technology and gadgets as most non-portable devices.

CONFLICT OF INTEREST

The authors confirm that this article content has no conflict of interest.

ACKNOWLEDGEMENTS

Declared none.

References

1
Hendee WR, Ritenour ER. Ultrasound waves In: Medical Imaging Physics. 4th ed. USA: Wiley-Liss 2002; pp. 303-29.
2
Graham DT, Cloke P. Ultrasound imaging. Appendix F In: Principles of radiological physics. 4th ed. Edinburgh: Churchill-Livingstone 2003; pp. 393-5.
3
Whaites E. Alternative and specialized imaging modalities In: Essentials of dental radiography and radiology. 4th ed. Edinburgh: Churchill-Livingstone 2007; pp. 223-41.
4
Carovac A, Smajlovic F, Junuzovic D. Application of ultrasound in medicine Acta Inform Med 2011; 19(3): 168-71.
5
Wakasugi-Sato N, Kodama M, Matsuo K, et al. Advanced clinical usefulness of ultrasonography for diseases in oral and maxillofacial regions 2010.
6
Dangore-Khasbage S, Degwekar SS, Bhowate RR, et al. Utility of color Doppler ultrasound in evaluating the status of cervical lymph nodes in oral cancer Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2009; 108(2): 255-63.
7
Joshi PS, Pol J, Sudesh AS. Ultrasonography - A diagnostic modality for oral and maxillofacial diseases Contemp Clin Dent 2014; 5(3): 345-51.
8
Kodama M, Khanal A, Habu M, et al. Ultrasonography for intraoperative determination of tumor thickness and resection margin in tongue carcinomas J Oral Maxillofac Surg 2010; 68(8): 1746-52.
9
Yuasa K, Kawazu T, Kunitake N, et al. Sonography for the detection of cervical lymph node metastases among patients with tongue cancer: criteria for early detection and assessment of follow-up examination intervals AJNR Am J Neuroradiol 2000; 21(6): 1127-32. [internet].
10
Wojtczak J, Bonadonna P. Pocket mobile smartphone system for the point-of-care submandibular ultrasonography Am J Emerg Med 2013; 31(3): 573-7.
11
Cornec D, Jousse-Joulin S, Marhadour T, et al. Salivary gland ultrasonography improves the diagnostic performance of the 2012 American College of Rheumatology classification criteria for Sjögren’s syndrome Rheumatology (Oxford) 2014; 53(9): 1604-7.
12
Poweski L, Drum M, Reader A, Nusstein J, Beck M, Chaudhry J. Role of ultrasonography in differentiating facial swellings of odontogenic origin J Endod 2014; 40(4): 495-8.
13
Assaf AT, Kahl-Nieke B, Feddersen J, Habermann CR. Is high-resolution ultrasonography suitable for the detection of temporomandibular joint involvement in children with juvenile idiopathic arthritis? Dentomaxillofac Radiol 2013; 42(3): 20110379.
14
Zengel P, Schrötzlmair F, Reichel C, Paprottka P, Clevert DA. Sonography: the leading diagnostic tool for diseases of the salivary glands Semin Ultrasound CT MR 2013; 34(3): 196-203.
15
Evans DH, Jensen JA, Nielsen MB. Ultrasonic colour Doppler imaging Interface Focus 2011; 1(4): 490-502.
16
Quiñones MA, Otto CM, Stoddard M, Waggoner A, Zoghbi WA. Recommendations for quantification of Doppler echocardiography: a report from the Doppler quantification task force of the nomenclature and standards committee of the American Society of Echocardiography J Am Soc Echocardiogr 2002; 15(2): 167-84.
17
Fenster A, Parraga G, Bax J. Three-dimensional ultrasound scanning Interface Focus 2011; 1(4): 503-19.
18
Mukhi PU, Mahindra UR. The use of ultrasonography in diagnosis and management of superficial fascial space infections 2012.
19
Pandey PK, Umarani M, Kotrashetti S, Baliga S. Evaluation of ultrasonography as a diagnostic tool in maxillofacial space infections J Oral Maxillofac Res 2012; 2(4): e4.
20
Tikku AP, Kumar S, Loomba K, Chandra A, Verma P, Aggarwal R. Use of ultrasound, color Doppler imaging and radiography to monitor periapical healing after endodontic surgery J Oral Sci 2010; 52(3): 411-6.
21
Lakshmi C, Sudhakara M, Ravikiran A, Sathish S, Bhavana SM. Evaluation of reliability of ultrasonographic parameters in differentiating benign and metastatic cervical group of lymph nodes 2014.
22
Poweski L, Drum M, Reader A, Nusstein J, Beck M, Chaudhry J. Role of ultrasonography in differentiating facial swellings of odontogenic origin J Endod 2014; 40(4): 495-8.
23
Singh GP, Dogra S, Kumari E. Ultrasonography: Maxillofacial applications Ann Dent Spec 2014; 2(3): 104-7.