RESEARCH ARTICLE
Impact of Cone Beam Computed Tomography Dose in Pre-Surgical Implant Analysis
Nabil EL Sahili1, *, Ibrahim Nasseh2, Antoine Berberi3, Sandra David-Tchouda3, 4, Sophie Thoret5, Thomas Fortin6, 7
Article Information
Identifiers and Pagination:
Year: 2018Volume: 12
Issue: Suppl-1, M6
First Page: 94
Last Page: 103
Publisher ID: TODENTJ-12-94
DOI: 10.2174/1874210601812010094
Article History:
Received Date: 17/11/2017Revision Received Date: 20/12/2017
Acceptance Date: 05/01/2018
Electronic publication date: 31/01/2018
Collection year: 2018

open-access license: This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 International Public License (CC-BY 4.0), a copy of which is available at: (https://creativecommons.org/licenses/by/4.0/legalcode). This license permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Abstract
Objectives:
Cone-Beam Computed Tomography (CBCT) produces vital information required for the accurate and prudent placement of dental implants. Lack of standardization between CBCT machines may result in unsafe patient exposure to harmful radiation; higher doses are not necessarily associated with improved image quality.
Aim:
The study aimed to assess the influence of low- and high-dose milliamperage settings on CBCT images for objective and subjective implant planning.
Methods:
Two dry skulls (4 hemi-maxillary segments of the maxilla and 4 hemi-maxillary segments of the mandible) were scanned under low (2 mA) and high (6.3 mA) dosage settings using a CBCT (Carestream CS 9300). Cross-sectional slices of both image qualities were evaluated by five expert clinicians, for image quality for implant planning and objective bone measurements.
Results:
There were no significant differences in bone measurements taken on high or low dose images (p > 0.05). In qualitative image assessments, assessment and image quality for almost all observers were independent of each other. For planning posterior mandibular implant placement, increased dosage improved concordance and kappa values between low and high dose images.
Conclusion:
Reduction in milliamperage did not affect diagnostic image quality for objective bone measurements and produced sufficient intra-rater reliability for qualitative assessment; therefore dose reduction can be achieved without compromising diagnostic decision- making.
Keywords: Cone-beam computed tomography, Implant-placement planning, Image quality, Radiation dose alteration, Posterior mandibular implant, Hemi Maxillary segments (HM).